Operation Operation
82 83
Online Help Shows Online-Help instead of the actual values in the menu
SET
pressure
, AUTO
easy
, AUTO
accurate
, Gradient, Favorites and
Setup.
Safety pressure If the system reaches the setting of the Safety Pressure
point, the vacuum control and ask rotation is stopped, the
evaporator is vented and the motor lift is raised (ML only).
Setting range is between 900 and 1200 mbar.
Heating Bath Media Allows switching between different heating bath media and
sets a limit to the maximum heating bath temperature for
the media for operator safety.
Max. Temp. Heating
Bath
Sets the limit of the maximum heating bath temperature
which can be set. The chosen bath media has an inuence on
the temperature range. Maximum set range between 20 and
210 °C.
Language
Sets language of the interface.
Bath Power at End
Sets the state of the bath after the process has ended. When
setting is “off” the bath is powered down after the process is
terminated.
Brightness Display
Changes brightness of the display.
Calibrate p-sensor
Opens sub-menu for calibration of the pressure sensor.
Error history
Shows the errors and history of occurrence for service data.
Reboot setup All settings of the Setup are put into the state of factory
setting when Reboot engaged.
Scale time axis Sets range of the visible time frame in the graphical “display
actual values”. Set range is between 1 and 120 minutes.
Setup Menu topology
Actual value display for all modes
Menu entry Value range
Vacuum
1 and 1200 mbar; default value: 1000 mbar
Vapor Vapor temperature measured at the vapor temperature
sensor.
Rotation
10 and 290 rpm in 1 rpm increments; default value: 100 rpm
Heating bath
20 °C and the max. temperature of the heating bath (to be
assigned in Setup), however, no more than 210 °C.
Softkey 1
Start Vacuum
Softkey 2 Start all
Stop all
Softkey 3
Max. Vacuum
Softkey 4
Stop Vacuum
SET
pressure
Mode
Menu entry Value range
Heating bath 20 °C and the max. temperature of the heating bath (to be
assigned in Setup), however, no more than 210 °C.
Rotation
10 and 290 rpm in 1 rpm increments; default value: 100 rpm
Vacuum
1 and 1200 mbar; default value: 1000 mbar
Hysteresis
1 to 50 mbar; default value: 10 mbar
Timer
1 to 1440 min; in 1 min; default value: 0 min; 0=off
Accept (OK)
Softkey 1
Store Favorite
Softkey 4
Esc